Skip to content

Public FDA record

Z-0197-2023: Beaver Arthro-Lok Pointed Tip Curve L 4mm Blade-an orthopedic blade, which may be used in various types o

Recall record Z-0197-2023 for Beaver Visitec International, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0197-2023
Company in FDA record
Beaver Visitec International, Inc.
FDA classification date
Nov 8, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect configuration of the blades in the package. The blades curve to the right instead of curving to the left. If used may cause cause unintended tissue damage during the procedure or if noticed prior to use may cause a delay in surgery

Source facts

Action
Beaver Visitec informed by a written letter notification coupled with an email on 9/30/22. Letter states reason for recall, health risk and action to take: 1. Immediately examine your inventory and quarantine product from these lots that are subject to recall. If you have further distributed this product, please notify your customers by initiating a sub-recall. Consider all potential users of this product in your user supply chain. You are encouraged to use a copy of this recall notification letter when contacting your customers. 2. Complete the enclosed Business Response Form (BRF) and indicate whether or not you have affected product to return. Please return the business response form to beavervisitec2765@sedgwick.com even if you no longer have product on hand. 3. Please enter the following information on the enclosed BRF: company name, lot number(s) and quantity(ies) of device(s) that are being returned. 4. If you have any questions regarding this Field Action, please contact BVI Sedgwick at 866-382-8606 or beavervisitec2765@sedgwick.com. If you have affected product, please return product with copy of BRF to: Return your product to SEDGWICK using the enclosed UPS pre-paid return label: Attn: Event 2765 Sedgwick 2670 Executive Drive, Suite A Indianapolis, IN 46241 A credit will be provided to consignees who have replied with confirmation of receiving affected product. Please direct any questions regarding credit to our customer service department by email ClaimsUS@bvimedical.com and include in the subject line: Master Case PIR 00371817.
Initiated
2022-09-30
Root cause
Under Investigation by firm
Distribution
FL, GA, IN, LA, MA, ME, VA Foreign: France, Japan, Netherlands, Switzerland, United Kingdom
Recall number
Z-0197-2023
Recall status
Open, Classified
CDRH record ID
196483
FDA classified
2022-11-08
Recall event number
91005

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗