It was found during a study, arcing might occur in the transformer of the high voltage generator, XTP-8100G and XTBP-8100G displaying an error message and interrupting fluoroscopy.
Source facts
Action
Consignees notification that a local TAMS service representative will schedule the implementation of the replacement transformer.
Initiated
2005-09-21
Root cause
Other
Terminated
2007-07-12
Distribution
CA, CO, KS, IL, LA, IN, ME, MI, NC, NJ, NY, OH, PA, TX
Recall number
Z-0198-06
Recall status
Terminated
CDRH record ID
42301
FDA classified
2005-11-24
Recall event number
33694