Open original FDA record ↗Recall record
- FDA identifier
- Z-0198-2013
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Nov 16, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.
Source facts
- Action
- GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue. Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard. Further questions please call (262) 513-4122.
- Initiated
- 2012-09-20
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2013-12-05
- Distribution
- Nationwide Distribution
- Recall number
- Z-0198-2013
- Recall status
- Terminated
- CDRH record ID
- 114302
- FDA classified
- 2012-11-16
- Recall event number
- 63597