Z-0198-2018: BD PhoenixTM AP, Catalog Number 448010 The BD Phoenix" AP instrument is designed for use with the BD Phoe

Recall record Z-0198-2018 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0198-2018
Company in FDA record
Becton Dickinson & Co.
FDA record creation date
Dec 6, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Potential unexpected movement of robot arm

Source facts

Action
BD Diagnostic Systems telephoned all customers on June 1, 2017 and sent an Urgent Product Recall letter dated May 31, 2017, and Response Forms to their customers via email. BD recently confirmed a manufacturing issue with the motor that drives movement of the robot arm. The issue may result in the inability of this motor to withstand the lateral forces encountered during normal operations. Should this occur, the instrument will indicate an E20 error-Pipettor step loss during operation and stop functioning. While there is no loss of patient sample, the samples would need to be manually processed until the instrument can be repaired. To limit any disruption in the workload, BD Service will contact customers to schedule the instrument replacement. Customers are advised to complete and return the response form via email to BDRC2@aol.com. Customers may continue to use the instrument until replacement is performed. Customers with questions may contact BD (Customer/Technical) Support at 1-800-638-8663. For questions regarding this recall call 410-316-4258.
Initiated
2017-06-01
Root cause
Process control
Terminated
2018-05-09
Distribution
Worldwide Distribution - US including CA, NC & TN Internationally: Canada
Recall number
Z-0198-2018
Recall status
Terminated
CDRH record ID
158667
FDA record created
2017-12-06
Recall event number
78102

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗