Open original FDA record ↗Recall record
- FDA identifier
- Z-0198-2019
- Company in FDA record
- Bovie Medical Corporation
- FDA record creation date
- Oct 19, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential risk associated with the PEEK tip of the J-Plasma Precise 360 Handpiece separating from the handpiece and retained in the tissue.
Source facts
- Action
- The firm, Bovie Medical, notified customers regarding the recall on or about 06/18/2018 via a letter entitled "URGENT: MEDICAL DEVICE RECALL". The letter alerted customers to discontinue to use the device with the affected catalog and lot numbers, to return the impacted handpieces to Bovie Medical, to consider using a device which did not contain the defective component, and how to receive a replacement device. Bovie Medical indicated the resolution to this issue will be a manufacturing and/or design change to the affected handpieces. The letter also stated to complete and return the response form included with the letter within ten(10) days by email to corrections@boviemed.com or by fax to the Regulatory Affairs Department at (727) 388-4697. Questions and comments can be directed to corrections@boviemed.com or call toll free 1-800-537-2790, Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time.
- Initiated
- 2018-06-18
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-04-08
- Distribution
- Worldwide Distribution: US (nationwide) to states of:: CA, CO, FL, ID, IL, MD, MO, NJ, NY, OH, OR, PA,, TX.; and countries of: Austria, Bahrain, Belgium, Bulgaria, Cyprus, Finland, Great Britain, Hungary, Ireland, Israel, Italy, Qatar, Scandinavia, Slovakia, Switzerland, Turkey, and United Arab Emirates.
- Recall number
- Z-0198-2019
- Recall status
- Terminated
- CDRH record ID
- 165712
- FDA record created
- 2018-10-19
- Recall event number
- 80544