Z-0198-2024: RX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040

Recall record Z-0198-2024 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0198-2024
Company in FDA record
Randox Laboratories Ltd.
FDA classification date
Oct 31, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Source facts

Action
Randox issued Medical Device Correction letter (REC687) on 9/13/23 to to US and PR customers on 9/12/23. Letter states reason for recall, health risk and action to take: Randox recommendations: " Auto rerun settings are disabled, see Appendix A for guidance for each Rx instrument. " Results with flag(<) reported as the bottom of the technical range. Sample Volume settings Low should not be applied. " Results with flag (>) reported as the top of the technical range. Sample Volume settings high and dilutions should not be applied. A manual dilution should be performed, and results multiplied by the dilution factor. The RX Parameter rerun settings will be updated with a parameter release in due course, the Instructions For Use will also be updated at this time. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation.
Initiated
2023-09-06
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution including Puerto Rico
Recall number
Z-0198-2024
Recall status
Open, Classified
CDRH record ID
203442
FDA classified
2023-10-31
Recall event number
93102

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗