Open original FDA record ↗Recall record
- FDA identifier
- Z-0199-2016
- Company in FDA record
- Interactive Motion Technologies, Inc.
- FDA classification date
- Nov 2, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unsupervised patient could potentially mis-use the device and remove or rotate a protective cover and expose a potential pinch hazard.
Source facts
- Action
- Interactive Motion Technology ( IMT) sent an Important Patient Safety Information letter dated March 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Summary of Actions to be taken For Hand Robot Cover B, firm recommends continued vigilance for both Adult and Pediatric Patients while in contact with this group of devices. For Hand Robot Covers A or C, at clinician discretion based on expected patient population, firm recommends continued use when appropriate of the IMT supplied Hypertonia Spring Assembly, or if not appropriate, insertion of the IMT Zero Force Spring Assembly. IMT will make this kit available for each Hand Robot upon request, or insert the Assembly when the Hand Robot is shipped to IMT using the special shipping container. If you have any questions, please contact your local IMT Representative or IMT Technical Services at support@interactive-motion.com.
- Initiated
- 2012-05-05
- Root cause
- Component design/selection
- Terminated
- 2017-07-13
- Distribution
- Worldwide Distribution - US Distribution to the states of : Texas and Tennessee., and to the countries of : China, Brazil, Hong Kong, Austria and South Korea.
- Recall number
- Z-0199-2016
- Recall status
- Terminated
- CDRH record ID
- 141071
- FDA classified
- 2015-11-02
- Recall event number
- 72326