Z-0199-2019: Nuclear Magnetic Resonance Imaging Systems that are indicated for use as a diagnostic imaging modality th

Recall record Z-0199-2019 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0199-2019
Company in FDA record
Canon Medical System, USA, INC.
FDA record creation date
Oct 22, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value. The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.

Source facts

Action
On 05/20/2018 the firm sent an Urgent Medical Device Correction letter to customers via United States Postal Service return receipt mail informing them that when a user used the respiratory gating with Steady Mode ON in Spin Echo sequence scan, the SAR value displayed on console was lower than the actual SAR value because there was a wrong estimating the RF pulses in the software. The SAR may exceed the upper limit, when a user scans a patient near the upper limit of the first level controlled operating mode in case of using the respiratory gating with Steady Mode ON in Spin Echo sequence scan. That may cause a burn of a patient. Applicable System/Model: -MEXL-1504/MEXL-1510: Vantage Titan -MEXL-1520: Vantage Elan Corrective Action: Modified system software to correct this issue will be installed on your system. When the software becomes available, your Canon service representative will contact your for an appointment to schedule the installation. Request to Customers: Until the modification is completed on your system, do not use Spin Echo sequences when using respiratory gating with Steady Mode ON. If you wish to use respiratory gating with Steady Mode ON, use Fast Spin Echo sequences. Please share this information with all users and reviewing radiologists as well as the clinical engineering or biomedical group at your facility. Customers are instructed to please complete the attached form and fax it to the toll-free number shown at the top of the form. This form can also be sent via email to RAffairs@us.medical.canon. If you have any questions, please feel free to contact Regulatory Affairs at (800) 421-1968 or your local service representative at (800) 521-1968.
Initiated
2018-05-30
Root cause
Device Design
Distribution
Domestic: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY. International: Algeria, Australia, Bangladesh, Brazil, Canada, China, Colombia, Congo, Dominican Republic, Egypt, Guinea, Honduras, India, Indonesia, Iran, Japan, Kenya, Korea, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Nigeria, Oman, Pakistan, Paraguay, Qatar, Saudi Arabia, Singapore, South Africa, Sudan, Taiwan, Thailand, Turkey, UAE, and Vietnam
Recall number
Z-0199-2019
Recall status
Open, Classified
CDRH record ID
167258
FDA record created
2018-10-22
Recall event number
81026

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗