Open original FDA record ↗Recall record
- FDA identifier
- Z-0199-2022
- Company in FDA record
- Bayer Medical Care, Inc.
- FDA record creation date
- Nov 3, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Labeling error: MEDRAD Mark 7 Arterion Injection System Syringes (ART 700 SYR) labeled as Twist & Go Syringes (TAG 150 SYR)
Source facts
- Action
- Bayer issued Urgent Medical Device Recall Letter (Recall ID: SA-21-RAD-02) dated 9/14/21 via email. Letter states reason for recall, health risk and action to take: Review your current inventory for affected Lot 8415317 and quarantine, as appropriate. 2. Complete the included response form and submit through email to CLSCRiCustomerLogistics@bayer.com or contact the Customer Logistics team at 724-940-7890. Please complete the Customer Response Form regardless of whether you have affected product or not. This action will assist us in tracking all affected product. 3. You will receive a Return Goods Authorization (RGA) number to return any affected Syringe kits remaining in your inventory for a credit. We can send new syringe kits with a separate purchase order, as necessary. 4. Place the affected product in a box and label the outside of the container with the RGA number in large, bold writing. Use the shipping label provided via email when you receive your RGA number. Please return affected product no later than November 30, 2021. 5. Upon Bayer receipt of returned product, you will receive a credit on your account. Please distribute this letter to other appropriate departments and personnel within your facility who may need to maintain awareness of this recall.
- Initiated
- 2021-09-14
- Root cause
- Packaging process control
- Terminated
- 2023-01-19
- Distribution
- Worldwide distribution - US Nationwide and the countries of AU, BY, DE, DK, GB, HR, NL, NZ, RU, SE, TH.
- Recall number
- Z-0199-2022
- Recall status
- Terminated
- CDRH record ID
- 189597
- FDA record created
- 2021-11-03
- Recall event number
- 88807