Scalpel and StatLock components packaged with specific lots of the HydroPICC catheter have a shorter expiration date than what is listed on the outer kit package.
Source facts
Action
Beginning February 24, 2021, the firm notified its affected consignees via Email.
Customers were informed that if product had already been used, then no action was required, because all components were still within their expiration date at the time, and no patient safety risk was identified. If product had not been used, customers were asked to refrain from using product until a replacement Scalpel and Statlock were provided.
Initiated
2021-02-24
Root cause
Process design
Terminated
2024-05-30
Distribution
Domestic distribution to NC, RI, NJ, and IL. No foreign distribution.
Recall number
Z-0199-2023
Recall status
Terminated
CDRH record ID
196380
FDA classified
2022-11-08
Recall event number
90981