Open original FDA record ↗Recall record
- FDA identifier
- Z-0199-2027
- Company in FDA record
- Philips North America
- FDA classification date
- Oct 16, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).
Source facts
- Action
- On september 16, 2026, URGENT Medical Device Correction, letters were sent to customers. Actions to be taken: 1. Direct clinical attendance of the patient during transport, when this issue is most likely to manifest, should be considered if not currently a part of the clinical workflow. In the event a persistent No Central Monit technical alarm is noted at the wireless device or a No Data technical alarm is noted at the central monitor; a restart of the wireless monitor may restore central monitoring when the patient is in known good wireless coverage area. 2. Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. Actions planned by Philips to correct the problem 3. Philips representative will contact you to arrange the software upgrade SH 2.0 AP 1.4GHz which will correct this issue. If you need any further information, please contact your local Philips representative, or call (800) 722-9377.
- Initiated
- 2026-09-16
- Root cause
- Software design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0199-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222860
- FDA classified
- 2026-10-16
- Recall event number
- 99908