Z-0199-2027: Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;

Recall record Z-0199-2027 for Philips North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0199-2027
Company in FDA record
Philips North America
FDA classification date
Oct 16, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).

Source facts

Action
On september 16, 2026, URGENT Medical Device Correction, letters were sent to customers. Actions to be taken: 1. Direct clinical attendance of the patient during transport, when this issue is most likely to manifest, should be considered if not currently a part of the clinical workflow. In the event a persistent No Central Monit technical alarm is noted at the wireless device or a No Data technical alarm is noted at the central monitor; a restart of the wireless monitor may restore central monitoring when the patient is in known good wireless coverage area. 2. Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. Actions planned by Philips to correct the problem 3. Philips representative will contact you to arrange the software upgrade SH 2.0 AP 1.4GHz which will correct this issue. If you need any further information, please contact your local Philips representative, or call (800) 722-9377.
Initiated
2026-09-16
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-0199-2027
Recall status
Open, Classified
CDRH record ID
222860
FDA classified
2026-10-16
Recall event number
99908

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗