Skip to content

Public FDA record

Z-0200-2007: Gambro Phoenix Hemodialysis Machine, Model numbers: 6022933700, 6023006700, 6022966700, Gambro Dasco S.p.

Recall record Z-0200-2007 for Gambro Renal Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0200-2007
Company in FDA record
Gambro Renal Products, Inc.
FDA classification date
Nov 21, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Defective Heat Exchanger/membrane; this may result in perforation of the membrane thus allowing the presence of infectious organisms and/or pathogens to pass from patient to patient.

Source facts

Action
Consignees were notified by fax on 05/12/2006 and by letter on 05/16/2006. Until it was determined that their Phoenix system did not have an affected heat exchanger, users were instructed to disinfect the hydraulics flow path after each patient treatment. Gambro personnel will inspect each system and replace any affected heat exchangers.
Initiated
2006-05-12
Root cause
Nonconforming Material/Component
Terminated
2012-01-03
Distribution
Worldwide Distribution, USA, including VA facilities in GA and NC. Foreign distribution to Canada, Korea and Mexico.
Recall number
Z-0200-2007
Recall status
Terminated
CDRH record ID
48227
FDA classified
2006-11-21
Recall event number
36233

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗