Defective Heat Exchanger/membrane; this may result in perforation of the membrane thus allowing the presence of infectious organisms and/or pathogens to pass from patient to patient.
Source facts
Action
Consignees were notified by fax on 05/12/2006 and by letter on 05/16/2006. Until it was determined that their Phoenix system did not have an affected heat exchanger, users were instructed to disinfect the hydraulics flow path after each patient treatment. Gambro personnel will inspect each system and replace any affected heat exchangers.
Initiated
2006-05-12
Root cause
Nonconforming Material/Component
Terminated
2012-01-03
Distribution
Worldwide Distribution, USA, including VA facilities in GA and NC. Foreign distribution to Canada, Korea and Mexico.
Recall number
Z-0200-2007
Recall status
Terminated
CDRH record ID
48227
FDA classified
2006-11-21
Recall event number
36233