Open original FDA record ↗Recall record
- FDA identifier
- Z-0201-06
- Company in FDA record
- Ferno-Washington Inc
- FDA classification date
- Nov 26, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm received complaints of the stretchers folding. A fold is when the stretcher lowers on its own to either the next position, or any of the other seven postions of the stretcher.
Source facts
- Action
- The firm went out to customers who had complained of a 'fold' and installed the kit on the stretcher that had experienced the fold and on all other PROflexx stretchers owned by the customer. These customers had received a recall letter, dated March 28, 2003. For customers that had not complained of the stretchers 'folding', the firm sent a letter dated June 24, 2003, to all distributors requesting the distributor provide the firm with a list of customers that had received the PROflex stretchers. Once the firm had a list of all customers, they sent them a letter offering them the kit as an enhancement that they could purchase to help reduce the possibility of false locking.
- Initiated
- 2003-03-28
- Root cause
- Other
- Terminated
- 2008-10-29
- Distribution
- The product was distributed nationwide to 27 states: TN, MA, VT, IL, KY, NY, SC, WI, TX, MO, NH, CO, GA, NC, CA, VA, IA, PA, CT, OH, NJ, ID, WV, AL, AR, ME and IN.
- Recall number
- Z-0201-06
- Recall status
- Terminated
- CDRH record ID
- 42904
- FDA classified
- 2005-11-26
- Recall event number
- 34010