Z-0201-2019: GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2.

Recall record Z-0201-2019 for GE Medical Systems, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0201-2019
Company in FDA record
GE Medical Systems, LLC
FDA classification date
Dec 28, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue. UPDATE 5/27/2016: Recall has been updated to include an additional product which includes the Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3.

Source facts

Action
Consignees were sent on 12/7/2015 a GE Healthcare "Urgent Medical Device Correction" GEHC Ref#60888 letter dated December 7, 2015. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers, Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. UPDATE: Consignees were sent a revised letter on May 2, 2016. Previously the letter only listed MR750W Surgical Suite Scanners. Because 3.0T 6 Channel Flex Coils are used with the MR750, MR750W, and 3.0T HDxT (including HD23), these devices have been added to the letter. There are no other changes to the letter. UPDATE 5/27/16: Consignees were sent a revised letter on May 19, 2016. The letter topic stated "De-rating of 3.0T 6 Channel Flex Coil when used on Discovery MR750 running DV24, DV25, or DV25.1 Application Software" . The letter described the Safety Issue, Safety Instruction, Affected Product details, Product Correction and Contact Information (same as previous letter).
Initiated
2015-12-07
Root cause
Nonconforming Material/Component
Terminated
2018-11-08
Distribution
Worldwide Distribution- US Distribution including the states of AZ, CA, CO, IL IN, MI, MN, NY, OH, OK, SD, TX, WI, VA, DE, FL, MD, ME, MO, NC, NH, OR, PA, TN, WI, DC, PR and the countries of CANADA, CHINA, CZECH REPUBLIC, ITALY, JAPAN, NETHERLANDS, NEW ZEALAND, PORTUGAL, ROMANIA, SOUTH AFRICA, SWITZERLAND, TURKEY, SWEDEN, AUSTRALIA, BRAZIL, CZECH REPUBLIC, EGYPT, FRANCE, GERMANY, ITALY, MEXICO, RUSSIA, SAUDI ARABIA, TAIWAN, UNITED KINGDOM, INDONESIA, ESTONIA, REPUBLIC OF KOREA, POLAND, KUWAIT, SPAIN.
Recall number
Z-0201-2019
Recall status
Terminated
CDRH record ID
149267
FDA classified
2015-12-28
Recall event number
72892

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗