Z-0201-2027: ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Opt

Recall record Z-0201-2027 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0201-2027
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Oct 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 9/9/26 was sent to customers. Return the completed customer response form to Beckman Coulter. Upon receipt of your form, we will coordinate a replacement of your rotor(s). If you have another compatible rotor available that is not included in the Serial Numbers listed above, discontinue use of the affected rotor(s) and use the alternative rotor until your replacement rotor is received. A description of the compatible rotors is available in the instrument instructions for use. If no alternative compatible rotor is available and continued use is necessary, you may continue to use your current rotor until you receive a replacement rotor; however, continued use carries a possibility of rotor breakage which may result in significant damage to materials (i.e., samples) being run in the centrifuge and the centrifuge instrument. Should you experience rotor breakage, please contact your service representative. Use personal protective equipment, and standard laboratory operating procedures for the handling of biohazardous materials when opening the centrifuge to prevent exposure to damaged materials (i.e., samples). Once your completed response form is received, a Beckman Coulter representative will coordinate a free-of-charge replacement order for you. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via electronic system. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact Customer Support Center. "From our website: http://www.beckman.com "By
Initiated
2026-09-21
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution.
Recall number
Z-0201-2027
Recall status
Open, Classified
CDRH record ID
222797
FDA classified
2026-10-21
Recall event number
99882

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗