Skip to content

Public FDA record

Z-0202-06: PROFlex Transporter, Model number 35P. The catalog number for the model 35P is PT3550.

Recall record Z-0202-06 for Ferno-Washington Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0202-06
Company in FDA record
Ferno-Washington Inc
FDA classification date
Nov 26, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The firm received complaints of the stretchers folding. A fold is when the stretcher lowers on its own to either the next position, or any of the other seven postions of the stretcher.

Source facts

Action
The firm went out to customers who had complained of a 'fold' and installed the kit on the stretcher that had experienced the fold and on all other PROflexx stretchers owned by the customer. These customers had received a recall letter, dated March 28, 2003. For customers that had not complained of the stretchers 'folding', the firm sent a letter dated June 24, 2003, to all distributors requesting the distributor provide the firm with a list of customers that had received the PROflex stretchers. Once the firm had a list of all customers, they sent them a letter offering them the kit as an enhancement that they could purchase to help reduce the possibility of false locking.
Initiated
2003-03-28
Root cause
Other
Terminated
2008-10-29
Distribution
The product was distributed nationwide to 27 states: TN, MA, VT, IL, KY, NY, SC, WI, TX, MO, NH, CO, GA, NC, CA, VA, IA, PA, CT, OH, NJ, ID, WV, AL, AR, ME and IN.
Recall number
Z-0202-06
Recall status
Terminated
CDRH record ID
42908
FDA classified
2005-11-26
Recall event number
34010

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗