Device does not produce a visual or audible impedance respiration APN alarm when an impedance respiration apnea event occurs.
Source facts
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated July 12, 2017, to all affected consignees via certified mail. Consignees were advised to enable "Resp Rate (Impedance)" alarms and set the low limits to the factory defaults (NICU: 20/minute; Adult ICU: 5/minute). Consignees were also advised to enable audible alarms under Alarm Priorities for RR (Impedance) high/low. GE will provide a correction free of charge once it is available. GE will contact consignees to arrange device corrections. Consignees were instructed to call 1-800-558-7044 or a local representative with any questions pertaining to this recall.
Initiated
2017-07-13
Root cause
Software Design Change
Terminated
2021-04-01
Distribution
Worldwide Distribution - US including NY, WI; Foreign: Australia, France, Germany.
Recall number
Z-0202-2018
Recall status
Terminated
CDRH record ID
157969
FDA record created
2017-12-06
Recall event number
77818