Open original FDA record ↗Recall record
- FDA identifier
- Z-0202-2027
- Company in FDA record
- Beckman Coulter, Inc.
- FDA classification date
- Oct 21, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
Source facts
- Action
- An URGENT MEDICAL DEVICE RECALL notification letter dated 9/9/26 was sent to customers. Return the completed customer response form to Beckman Coulter. Upon receipt of your form, we will coordinate a replacement of your rotor(s). If you have another compatible rotor available that is not included in the Serial Numbers listed above, discontinue use of the affected rotor(s) and use the alternative rotor until your replacement rotor is received. A description of the compatible rotors is available in the instrument instructions for use. If no alternative compatible rotor is available and continued use is necessary, you may continue to use your current rotor until you receive a replacement rotor; however, continued use carries a possibility of rotor breakage which may result in significant damage to materials (i.e., samples) being run in the centrifuge and the centrifuge instrument. Should you experience rotor breakage, please contact your service representative. Use personal protective equipment, and standard laboratory operating procedures for the handling of biohazardous materials when opening the centrifuge to prevent exposure to damaged materials (i.e., samples). Once your completed response form is received, a Beckman Coulter representative will coordinate a free-of-charge replacement order for you. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via electronic system. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact Customer Support Center. "From our website: http://www.beckman.com "By
- Initiated
- 2026-09-21
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0202-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222799
- FDA classified
- 2026-10-21
- Recall event number
- 99882