Z-0203-2015: BrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates wi

Recall record Z-0203-2015 for Philips Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0203-2015
Company in FDA record
Philips Medical Systems, Inc.
FDA classification date
Dec 12, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered the CT portion of the scan was interrupted ( stopping the CT exposure at the time of interruption) and did not complete successfully.

Source facts

Action
Philips sent an Important Electronic Product Radiation Warning letter dated November 4, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken by the customer: Please be aware that if this condition occurs, users will not be able to detect the situation prior to the interruption of the CT acquisition. However, if the INFORMATION message appears during a clinical study indicating to the operator that the scan did not complete successfully, Philips advises users to understand the following available options, and use their clinical judgment to determine which option best suits their clinical needs: The operator chooses not to rescan the patient: Continue with the SPECT portion of the of the scan and use only the portion of the segments that were acquired prior to the interruption for interpretation; or The operator chooses to rescan the patient: Retry the interrupted segment acquisition. Philips Healthcare is initiating a corrective action consisting of : Distribution of this Field Safety Notice 88200485_486, Conducting the appropriate field safety correction will address the above issue that has been identified. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center (1-800-722-9377: follow the prompts).
Initiated
2013-10-23
Root cause
Radiation Control for Health and Safety Act
Terminated
2021-03-01
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AZ, CA, CO, DC, DE, FL, GA, IL, MA, MD, MN, MO, MS, MT, NC, NJ, NY, OH, OK, OR, PA, TX, VT, WA and WI., and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Finland, France, Gabon, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom and Viet Nam.
Recall number
Z-0203-2015
Recall status
Terminated
CDRH record ID
130194
FDA classified
2014-12-12
Recall event number
69304

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗