Skip to content

Public FDA record

Z-0203-2018: Fukuda Denshi patient monitor model DS-8100M and DS-8100N Product Usage: Use of the Fukuda Denshi DynaSc

Recall record Z-0203-2018 for Fukuda Denshi Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0203-2018
Company in FDA record
Fukuda Denshi Co., Ltd.
FDA record creation date
Dec 6, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

For the Fukuda Denshi patient monitor model DS-8100, in less than 0.003% (17) of the 6227 devices sold worldwide, a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term battery backup).

Source facts

Action
The recalling firm will notify affected consignees via Field Correction Notification letter. The letter identified the affected product, problem and actions to be taken. A Customer Confirmation Sheet to be filled out by end user and faxed or emailed back to Fukuda Denshi USA, Inc will also be included. For questions call 206-305-4460.
Initiated
2017-10-16
Root cause
Component design/selection
Distribution
Worldwide Distribution - US Nationwide and in the countries of UK, and Japan and via distributors in the other countries
Recall number
Z-0203-2018
Recall status
Completed
CDRH record ID
159575
FDA record created
2017-12-06
Recall event number
78441

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.