For the Fukuda Denshi patient monitor model DS-8100, in less than 0.003% (17) of the 6227 devices sold worldwide, a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term battery backup).
Source facts
Action
The recalling firm will notify affected consignees via Field Correction Notification letter. The letter identified the affected product, problem and actions to be taken. A Customer Confirmation Sheet to be filled out by end user and faxed or emailed back to Fukuda Denshi USA, Inc will also be included. For questions call 206-305-4460.
Initiated
2017-10-16
Root cause
Component design/selection
Distribution
Worldwide Distribution - US Nationwide and in the countries of UK, and Japan and via distributors in the other countries
Recall number
Z-0203-2018
Recall status
Completed
CDRH record ID
159575
FDA record created
2017-12-06
Recall event number
78441