Z-0203-2019: IM Abbott Alinity ci-series (abdicqii) accessory driver v8.00.0001 software product that can receive info

Recall record Z-0203-2019 for Data Innovations, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0203-2019
Company in FDA record
Data Innovations, LLC
FDA record creation date
Oct 22, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction.

Source facts

Action
On August 29, 2018, the firm began notifying affected customers via an Important Medical Device Recall/Field Safety Notice Information letter. The communication advised customers of the issue with Abbott Diagnostics Alinity ci-series (abdicqii) instrument not processing QC results if "Hold Results Until Complete" is enabled. Customers are advised to do the following: 1. Apply the updated v8.00.0002 version of Instrument Manager driver to correct this issue. The updated version is available around September 7, 2018. 2. Use the following workaround: Disable the "Hold Results Until Complete" option in Instrument Manager (driver configuration). If Hold Results Until Complete logic is required, there is no workaround; discontinue use until updated driver is installed. 3. Distribute notification on to anyone in your organization who needs to be aware. 4. Complete the customer response form and return to recalling firm. If you have any questions regarding this recall, please contact Support Email USA: northamerica-support@datainnovations.com, or Support Phone: +1(802) 658-1955.
Initiated
2018-08-29
Root cause
Software design
Terminated
2019-02-05
Distribution
Distributed to OK. Foreign distribution to Argentina, Hong Kong, Malaysia, Thailand, Germany, Jordan, Australia, and Switzerland.
Recall number
Z-0203-2019
Recall status
Terminated
CDRH record ID
167216
FDA record created
2018-10-22
Recall event number
81005

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗