Z-0203-2022: GM Helix Acqua Implant 3.5x10, Article No. 140.944

Recall record Z-0203-2022 for Jjgc Industria E Comercio De Materials Dentarios Sa. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0203-2022
Company in FDA record
Jjgc Industria E Comercio De Materials Dentarios Sa
FDA record creation date
Nov 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

3.5x11.5mm implants were mislabeled as 3.5x10mm implants. In borderline cases where the bone quality is low and the bone ridge has a maximum indication for the implant height (10mm), if the incorrect implant is installed, it could result in damage to the nerve or adjacent bone structures/cavities.

Source facts

Action
The firm initially contacted affected customers by phone on October 11, 2021. Customers were instructed to immediately stop using the product and to place all impacted product in quarantine. The firm then followed up with a confirmation letter sent by UPS on October 22, 2021. Customers were instructed to identify and segregate the affected units, open or closed, in their stock. Product is to be returned to the manufacturer.' If product has been installed and the patient reports an issue (such as numbness, pain, or other symptoms), please notify the distributor immediately to receive instructions on how to proceed.
Initiated
2021-10-11
Root cause
Under Investigation by firm
Terminated
2024-01-16
Distribution
US Nationwide distribution.
Recall number
Z-0203-2022
Recall status
Terminated
CDRH record ID
190005
FDA record created
2021-11-04
Recall event number
88917

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗