Open original FDA record ↗Recall record
- FDA identifier
- Z-0203-2027
- Company in FDA record
- Medtronic MiniMed, Inc.
- FDA classification date
- Oct 21, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).
Source facts
- Action
- On 9/11/2026, correction notices were mailed to the health care professional, and customer asking them to do the following: The Customer was asked to do the following: Immediately contact your Healthcare Provider (HCP) to confirm your insulin settings values. Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/FA1587 The health care professional was asked to do the following: If you provided diabetes treatment recommendations to patients based on custom Total Daily Dose(TDD) or long-acting insulin values entered into the CareLin Clinic Settings Calculator between: June 04, 2026, to June 08, 2026, please revisit those recommendations utilizing patients' individual CareLink reports. Continue to follow up with those patients pursuant to your standard of care. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.
- Initiated
- 2026-09-11
- Root cause
- Software design
- Distribution
- US Nationwide distribution in the state of KS.
- Recall number
- Z-0203-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222434
- FDA classified
- 2026-10-21
- Recall event number
- 99765