Z-0204-2020: Atellica IM 1300 Analyzer, (SW) version V1.19.0 and V1.19.2 Siemens Material Number (SMN): 11066001

Recall record Z-0204-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0204-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Oct 24, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Issues Scanning New Reagent Lot Test Definition on the Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer SW V 1.19.0 and 1.19.2

Source facts

Action
Siemens issued an Urgent Medical Device Correction (UMDC) to US customers via FedEx on August 15, 2019. The UFSN was electronically issued to all OUS Siemens sites, with affected customers, for distribution, as well as regional reporting to their respective local competent authority on August 16, 2019. The UMDC/UFSN explain the behavior(s) to users and provide actions to be taken to correct the behavior. Customers are requested to return the Effectiveness Check Form, in the UMDC/UFSN via fax or email. Siemens Healthcare Diagnostics Inc. has released an updated software (SW) version (V) 1.20.0, which corrects the reported behavior(s). A Mandatory Update (MU) campaign has been launched to install the new SW V on all systems, whether or not they are impacted by these behaviors.
Initiated
2019-08-15
Root cause
Under Investigation by firm
Terminated
2022-05-05
Distribution
Nationwide Foreign: Argentina Australia Bangladesh Belgium Brazil Colombia Dominican Republic Finland France Germany Greece Italy Japan Latvia Mexico Netherlands P.R. China Republic Korea Singapore Spain Switzerland Thailand United Kingdom Vietnam
Recall number
Z-0204-2020
Recall status
Terminated
CDRH record ID
176029
FDA record created
2019-10-24
Recall event number
83655

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗