Open original FDA record ↗Recall record
- FDA identifier
- Z-0205-2013
- Company in FDA record
- Ormco/Sybronendo
- FDA classification date
- Nov 5, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130C. The correct sterilization temperature should be listed as 132C and/or 270F.
Source facts
- Action
- "Urgent - Medical Device Correction" letters dated 12/12/2011 were sent via USPS 1st class mail to all customers and again on 2/21/2012 for recall expansion. The device issue was described and recommended actions were provided. Customers were instructed to acknowledge and complete the Return Form. Questions were referred to (800) 854-1741.
- Initiated
- 2011-12-12
- Root cause
- Error in labeling
- Terminated
- 2012-11-05
- Distribution
- Worldwide distribution, including Nationwide (USA) and the countries of Canada, Austria, Belgium, Bulgaria, Switzerland, Germany, Denmark, France, Estonia, Spain, Finland, Iceland, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Luxembourg, Netherlands, Monaco, Poland, Romania, Serbia, Slovenia, Turkey, Jersey, and Martinique.
- Recall number
- Z-0205-2013
- Recall status
- Terminated
- CDRH record ID
- 113714
- FDA classified
- 2012-11-05
- Recall event number
- 63412