Z-0205-2016: Arthroscopic Energy 50 Probe with Suction, Catalog number: AES-50S Peoduct Usage: For resection, ablatio

Recall record Z-0205-2016 for ConMed Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0205-2016
Company in FDA record
ConMed Corporation
FDA classification date
Nov 3, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Probe programming error. The incorrect program can generate more heat at the surgical site, resulting in an elevated probability of tissue damage.

Source facts

Action
On June 17, 2015 Con Med Corporation distributed Urgent Device Recall Notification Letters and Recall Reply forms to their customers via courier service. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory for any of the affected lot numbers referenced on the recall notice. If any of the recalled product are in stock, customers should discontinue the use or distribution of these items. If product was further distributed, customers should notify those individuals of this recall action. Customers with questions can call Gregory Connell at 727-399-5276 or email AES50S@conmed.com. Customers should also complete and return the Recall Reply Form by mail to the address listed on the recall notice.
Initiated
2015-06-17
Root cause
Software design (manufacturing process)
Terminated
2017-04-11
Distribution
Worldwide - US distribution in the states of FL, OH, and the country of Spain.
Recall number
Z-0205-2016
Recall status
Terminated
CDRH record ID
138104
FDA classified
2015-11-03
Recall event number
71518

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗