Open original FDA record ↗Recall record
- FDA identifier
- Z-0205-2027
- Company in FDA record
- TECHNO-PATH MANUFACTURING LTD.
- FDA classification date
- Oct 22, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Source facts
- Action
- On August 14, 2026, Techno Path Manufacturing Ltd has issued a medical device correction to consignees via email. Techno Path Manufacturing Ltd informed consignees to please verify their inventory against the affected product codes and lot numbers. Techno Path Manufacturing Ltd. has informed the consignees to take the following actions: 1. Immediately place all affected inventory on hold and prevent further distribution. 2. Quarantine all affected product currently under your control. 3. Confirm that there has been no distribution of the impacted lots to customers. 4. Confirm receipt of this notification and provide inventory and distribution details to Technopath Manufacturing Ltd. within 3 business days. 5. Arrange return or destruction of affected inventory in accordance with instructions provided by Technopath Manufacturing Ltd.
- Initiated
- 2026-08-14
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the state of Massachusetts.
- Recall number
- Z-0205-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222301
- FDA classified
- 2026-10-22
- Recall event number
- 99724