Skip to content

Public FDA record

Z-0206-2024: RayCare 5B and 6A including service packs- An oncology information system used to support workflows and s

Recall record Z-0206-2024 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0206-2024
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA classification date
Nov 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.

Source facts

Action
Raysearch issued Field Safety Notice, Medical Device Correction #121600 via email on 9/12/23. Letter states reason for recall, health risk and action to take: Educate staff and all users to open configurable documents from the Documents workspace to avoid the issue. " Review all configured documents and consider reconfiguration in a way that will not trigger the issue. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. Solution This issue will be resolved in the next version of RayCare, scheduled for market release in December 2023 (RayCare 2024A) (subject to market clearance in some markets). If customers wish to continue using versions of RayCare affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact quality@raysearchlabs.com.
Initiated
2023-09-12
Root cause
Under Investigation by firm
Distribution
MD, TN
Recall number
Z-0206-2024
Recall status
Open, Classified
CDRH record ID
203529
FDA classified
2023-11-01
Recall event number
93139

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗