Z-0206-2027: Techno-Path Multichem AE Unassayed Level 2 REF: AE622X

Recall record Z-0206-2027 for TECHNO-PATH MANUFACTURING LTD.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0206-2027
Company in FDA record
TECHNO-PATH MANUFACTURING LTD.
FDA classification date
Oct 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

Source facts

Action
On August 14, 2026, Techno Path Manufacturing Ltd has issued a medical device correction to consignees via email. Techno Path Manufacturing Ltd informed consignees to please verify their inventory against the affected product codes and lot numbers. Techno Path Manufacturing Ltd. has informed the consignees to take the following actions: 1. Immediately place all affected inventory on hold and prevent further distribution. 2. Quarantine all affected product currently under your control. 3. Confirm that there has been no distribution of the impacted lots to customers. 4. Confirm receipt of this notification and provide inventory and distribution details to Technopath Manufacturing Ltd. within 3 business days. 5. Arrange return or destruction of affected inventory in accordance with instructions provided by Technopath Manufacturing Ltd.
Initiated
2026-08-14
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the state of Massachusetts.
Recall number
Z-0206-2027
Recall status
Open, Classified
CDRH record ID
222305
FDA classified
2026-10-22
Recall event number
99724

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗