Open original FDA record ↗Recall record
- FDA identifier
- Z-0207-2012
- Company in FDA record
- Lifewatch Services Inc
- FDA classification date
- Nov 17, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
LifeWatch received complaints of minor shocks and burns during use of the LifeStar ACT systems with shorts in the leads.
Source facts
- Action
- The firm, LifeWatch, sent an " Urgent Device Recall" letter dated September 27, 2011 to the affected customers. The letter describes the product, problem and action to be taken. The customers were instructed to immediately refrain from providing these devices to their customers; stop using the affected sensors, return their entire kit in the pre-paid mailer, and use the new device kit included with the recall letter. In addition, the letter informed the users that the replacement device has upgraded software and firmware that allows for detection of shorts in the cable, and if detected, cuts off current flow to the patient to prevent minor shocks and burns. If you have any questions, please call (847) 813-4258.
- Initiated
- 2011-09-26
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-05-15
- Distribution
- Nationwide distribution.
- Recall number
- Z-0207-2012
- Recall status
- Terminated
- CDRH record ID
- 104184
- FDA classified
- 2011-11-17
- Recall event number
- 59990