Z-0207-2019: Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable I

Recall record Z-0207-2019 for Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0207-2019
Company in FDA record
Howmedica Osteonics Corp.
FDA record creation date
Oct 23, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space via the impaction method and the implant pivots unexpectedly, the implant potentially may unintentionally contact sensitive anatomy. Additionally, the user may potentially experience difficulty positioning the implant within the disc space due to the lack of rigidity. A third potential hazard is associated with the inserter not unlocking at the articulating pivot head during implant positioning within the disc space. Once the implant is in the prepared disc space and contacts the ventral annulus, the surgeon unlocks the articulating pivot head and positions the implant perpendicular to the spinous process. The surgeon may then elect to lock/unlock the articulating pivot head for desired final positioning of the implant. If the inserter does not unlock during these final implant positioning steps, the user may potentially implant the cage in a suboptimal trajectory/placement.

Source facts

Action
On September 12, 2018, the firm notified its sales personnel of the recall via a Product Correction letter. The letter identified the affected product, problem and actions to be taken. Applicable Stryker personnel that are involved in the maintenance or preparation of the Tritanium TL Inserter have been identified to receive notification of this correction to assure that the instrument is lubricated adequately.
Initiated
2018-09-12
Root cause
Under Investigation by firm
Terminated
2020-07-16
Distribution
US Nationwide Distribution.
Recall number
Z-0207-2019
Recall status
Terminated
CDRH record ID
167441
FDA record created
2018-10-23
Recall event number
81096

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗