Skip to content

Public FDA record

Z-0208-2012: ACT Cellular Phone Monitor; a component of the ACT Ambulatory Cardiac Telemetry; an automatically activat

Recall record Z-0208-2012 for Lifewatch Services Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0208-2012
Company in FDA record
Lifewatch Services Inc
FDA classification date
Nov 17, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

LifeWatch received complaints of minor shocks and burns during use of the LifeStar ACT systems with shorts in the leads.

Source facts

Action
The firm, LifeWatch, sent an " Urgent Device Recall" letter dated September 27, 2011 to the affected customers. The letter describes the product, problem and action to be taken. The customers were instructed to immediately refrain from providing these devices to their customers; stop using the affected sensors, return their entire kit in the pre-paid mailer, and use the new device kit included with the recall letter. In addition, the letter informed the users that the replacement device has upgraded software and firmware that allows for detection of shorts in the cable, and if detected, cuts off current flow to the patient to prevent minor shocks and burns. If you have any questions, please call (847) 813-4258.
Initiated
2011-09-26
Root cause
Nonconforming Material/Component
Terminated
2012-05-15
Distribution
Nationwide distribution.
Recall number
Z-0208-2012
Recall status
Terminated
CDRH record ID
104235
FDA classified
2011-11-17
Recall event number
59990

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗