Open original FDA record ↗Recall record
- FDA identifier
- Z-0208-2013
- Company in FDA record
- Sechrist Industries Inc
- FDA classification date
- Nov 6, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Sechrist conducted a field upgrade on the hyperbaulic chambers equipped with Electronic Control Modules in order to correct a software bug that could potentially affect the performance of the chamber.
Source facts
- Action
- Sechrist Industries contacted consignees by phone to notify them of a software upgarde and to schedule appointments to perform the required field upgrades. An approximate summary of the phone notification is provided below. "I am contacting you to notify you of a field upgrade we are performing on our hyperbaric chambers equipped with Electronic Control Modules in order to correct a software bug that could potentially affect the performance of your chamber. The upgrade includes the installation of a new motherboard with revised software, a reset switch, and the exchange of other control panel components. The process will take approximately 2.5 hours and will be performed at no charge to you." For questions regarding this recall call 714-579-8400.
- Initiated
- 2008-08-28
- Root cause
- Software design
- Terminated
- 2012-11-06
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Bulgaria, and Canada
- Recall number
- Z-0208-2013
- Recall status
- Terminated
- CDRH record ID
- 84024
- FDA classified
- 2012-11-06
- Recall event number
- 52796