Z-0208-2025: Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation

Recall record Z-0208-2025 for Percussionaire Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0208-2025
Company in FDA record
Percussionaire Corporation
FDA classification date
Oct 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.

Source facts

Action
On 10/03/2024, the firm emailed "Urgent Medical Device Recall" Letter informing customer that Percussionaire became aware of discoloration of fluid collecting in the Phasitron breathing circuit after extended Intrapulmonary Percussive Ventilation (IPV) therapy. I was determined that there is a reaction within the nebulizer cup between the combinations of saline mixed with albuterol and/or N-acetylcysteine, and an inadvertent nickel-coating on the spring. The reaction can aerosolize the nickel under normal clinical use. Customer are instructed to: " Complete and Return Acknowledgement form (see Appendix 1) after reviewing and implementing the requested actions below by October 18, 2024. " Return all impacted products " Until unaffected or replacement product is available, affected devices should not be used under the following conditions (Appendix 3): o For the delivery of medications o For use longer than 14 days when using normal saline o On patients with an inflammatory state or with a known nickel allergy o Immediately stop usage if any unexpected symptoms are encountered " Additionally, it is recommend to rinsing the Phasitron 5 after each use with sterile water and allowing it to air dry per the IFU. Percussionaire will: Replace the affected product and exchange with new, unaffected product. Until replacement or unaffected product is available, Customers have been advised to use the product only with saline and only under restricted conditions For questions/assistance - email FSCA@sentec.com
Initiated
2024-09-16
Root cause
Process control
Distribution
U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI, O.U.S.: Canada, Switzerland (all products shipped OUS remained in Sentec AG's (distributor, parent company of Percussionaire) control, were quarantined, and did not leave the warehouse).
Recall number
Z-0208-2025
Recall status
Open, Classified
CDRH record ID
210266
FDA classified
2024-10-30
Recall event number
95436

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗