Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.
Source facts
Action
Abbott issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 09/28/2023 via FedEx 2nd day delivery. The notice explained the problem, risk, and requested return of the device for inspection. Those devices which pass, will be returned. Should your device(s) fail the inspection, your local Abbott Representative can advise the consignee of alternate commercially available products.
Initiated
2023-09-28
Root cause
Process change control
Distribution
US, Taiwan, and Colombia
Recall number
Z-0209-2024
Recall status
Open, Classified
CDRH record ID
203550
FDA classified
2023-11-02
Recall event number
93152