Open original FDA record ↗Recall record
- FDA identifier
- Z-0210-2015
- Company in FDA record
- Fisher & Paykel Healthcare, Ltd.
- FDA classification date
- Nov 13, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.
Source facts
- Action
- Fisher & Paykel Healthcare sent an Urgent Medical Device Recall letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Step 1: Identify any affected Infant Warmers in your inventory by checking the REF and SERIAL number on the device label (see examples in Figures 1 and 2). Step 2: Identify any affected spare parts in your inventory by checking the REF and LOT number on the parts (see example in Figure 3). Step 3: Check your Infant Warmer service records to determine if affected spare parts were installed on any other Infant Warmers. If you are unable to identify specific Infant Warmers that may have had affected spare parts installed, then please contact FPH at the contact details below. Step 4: Replacement of Affected Nut. - Remove the affected Infant Warmer from use. - Switch the warmer off and disconnect it from the mains power supply. - Support the heater head during disassembly to prevent it from falling. - Follow Steps A to F in Figure 4 below to replace the nut. - Ensure the pivot washer remains in place when installing the new nut. - Ensure that the heater head is secure, can rotate smoothly, and the detent locates the head in the central position. - Reconnect the mains power supply, switch the warmer on, and ensure there are no error codes or alarms before returning to use. Step 5: Complete the Urgent Medical Device Recall Response Form and return it to your FPH Representative
- Initiated
- 2014-11-06
- Root cause
- Manufacturing material removal
- Terminated
- 2015-04-27
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Bekgium, Bermuda, Chile, China, Croatia, Dominican Republic, Estonia, Finland, France, Germany, Great Britain, Greece, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Mexico. Morocco, New Zealand, Nigeria, Northern Ireland, Norway, Pakistan, Palestinian Territory, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, singapore, Slovenia, South Africa, Spain, Sri Lanka, Sultanate of Oman, Syria, Taiwan, Thailand, The Netherlands, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam, and West Bank.
- Recall number
- Z-0210-2015
- Recall status
- Terminated
- CDRH record ID
- 131307
- FDA classified
- 2014-11-13
- Recall event number
- 69668