Skip to content

Public FDA record

Z-0210-2016: AMSCO 400 and AMSCO C Small Steam Sterilizers

Recall record Z-0210-2016 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0210-2016
Company in FDA record
Steris Corporation
FDA classification date
Nov 4, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

In AMSCO Small Steam Sterilizers equipped with the CS-iQ feature, the sterilizer software inadvertently prevents the capability to operate the foot pedal, which allows the Operator to open and close the sterilizer door.

Source facts

Action
On 9/21/2015 the firm sent Recall Correction Notices to their customers.
Initiated
2015-09-03
Root cause
Software design
Terminated
2016-01-22
Distribution
US nationwide distribution.
Recall number
Z-0210-2016
Recall status
Terminated
CDRH record ID
140236
FDA classified
2015-11-04
Recall event number
72217

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗