Open original FDA record ↗Recall record
- FDA identifier
- Z-0210-2018
- Company in FDA record
- Biosense Webster, Inc.
- FDA record creation date
- Dec 7, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets.
Source facts
- Action
- All new U.S. and OUS customers, who have purchased the SMARTABLATE" Generator system since the launch of the original field action in March 2015 to the time the IFU was updated in November 2015, will be notified. The previously distributed Field Safety Notification Letter for the SMARTABLATE" Generator (Part Numbers# M490006 and M490007) and customer Acknowledgement Form will be sent to these new customers. Instructions included review the Field Safety Notice carefully, provide the notice to appropriate personnel in facility, ensure the SMARTABLATE System Foot Pedal is not located within the Stereotaxis RMT Lab, complete and return the Acknowledgement Forms, and maintain awareness of the notice.
- Initiated
- 2017-06-14
- Root cause
- Labeling design
- Terminated
- 2021-01-11
- Distribution
- Distributed to the following states: CA, CO, GA, IL, IN, MI, NM, NY, OH, PA, TN, TX, UT, WA, WI. Product was also distributed to the following countries: Germany, France, Austria, Sweden Italy, Belgium, Denmark, and Singapore.
- Recall number
- Z-0210-2018
- Recall status
- Terminated
- CDRH record ID
- 156581
- FDA record created
- 2017-12-07
- Recall event number
- 77575