Z-0210-2022: Covidien Emprint Ablation Catheter with Thermosphere Technology Bronchoscopic Ablation, CFN CA108L1

Recall record Z-0210-2022 for Covidien Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0210-2022
Company in FDA record
Covidien Llc
FDA record creation date
Nov 5, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm identified an error in the device temperature monitoring circuitry which is designed to interrupt energy delivery in the event of a primary cooling system failure. Use of an affected device in conjunction with a primary cooling system failure could lead to the applicator tip overheating and associated heat damage which could result in an allergic/toxic reaction, unintended tissue ablation, tissue burn, and/or a delay of treatment while an alternate device is located.

Source facts

Action
Beginning September 29, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were asked to immediately quarantine and discontinue use of the affected products. Customers should return affected product to the firm. If product has been further distributed, please forward the recall information. Customers should complete and return the accompanying response form even if they do not have product on hand.
Initiated
2021-09-29
Root cause
Process control
Terminated
2025-02-13
Distribution
No domestic distribution. Foreign distribution to Hong Kong and UK.
Recall number
Z-0210-2022
Recall status
Terminated
CDRH record ID
189622
FDA record created
2021-11-05
Recall event number
88817

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗