Open original FDA record ↗Recall record
- FDA identifier
- Z-0210-2022
- Company in FDA record
- Covidien Llc
- FDA record creation date
- Nov 5, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm identified an error in the device temperature monitoring circuitry which is designed to interrupt energy delivery in the event of a primary cooling system failure. Use of an affected device in conjunction with a primary cooling system failure could lead to the applicator tip overheating and associated heat damage which could result in an allergic/toxic reaction, unintended tissue ablation, tissue burn, and/or a delay of treatment while an alternate device is located.
Source facts
- Action
- Beginning September 29, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were asked to immediately quarantine and discontinue use of the affected products. Customers should return affected product to the firm. If product has been further distributed, please forward the recall information. Customers should complete and return the accompanying response form even if they do not have product on hand.
- Initiated
- 2021-09-29
- Root cause
- Process control
- Terminated
- 2025-02-13
- Distribution
- No domestic distribution. Foreign distribution to Hong Kong and UK.
- Recall number
- Z-0210-2022
- Recall status
- Terminated
- CDRH record ID
- 189622
- FDA record created
- 2021-11-05
- Recall event number
- 88817