Open original FDA record ↗Recall record
- FDA identifier
- Z-0211-2008
- Company in FDA record
- Teleflex Creek Dr
- FDA classification date
- Dec 28, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Decreased in flow output; through the nebulizer -(when used at higher oxygen concentration settings, this decrease in flow, may lower the oxygen concentration inspired by the patient)
Source facts
- Action
- Customers were notified by an Urgent Medical Device Recall letter on/about 08/22/2007. They were instructed to return the affected products to Teleflex Medical or acknowledge that the products no longer exist. In addition, each were instructed to forward the letter if they have further distributed into institutional settings or to homecare patients and retrieve relevant product from those locations. This recall is to the consumer/user level via sub-recall from distributors or direct accounts as per the Recalling firm.
- Initiated
- 2007-08-22
- Root cause
- Release of Material/Component prior to receiving test results
- Terminated
- 2012-08-13
- Distribution
- Worldwide: USA, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, France, Germany, Guatemala, India, Mexico, Panama, Peru, Philippines, South Korea, Thailand, United Arab Emirates, and The United Kingdom.
- Recall number
- Z-0211-2008
- Recall status
- Terminated
- CDRH record ID
- 65454
- FDA classified
- 2007-12-28
- Recall event number
- 45426