Skip to content

Public FDA record

Z-0211-2012: IMPAX Cardiovascular Results Management (RM) Software Versions RM 2.04.37.04 to RM 7.8 SU2 The IMPAX Car

Recall record Z-0211-2012 for AGFA Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0211-2012
Company in FDA record
AGFA Corp.
FDA classification date
Nov 18, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Improper merging of IMPAX Cardiovascular Results Management (RM) reports causes the signed report to become unavailable.

Source facts

Action
AGFA Healthcare sent an "URGENT FIELD SAFETY NOTICE" letter dated May 27, 2011, via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to carry out a merge post finalization of an RM reports. If it is necessary to modify a finalized report, AGFA recommends that they make the modification through one of the following actions: Amend the report through the RM interface, or Call AGFA Service to change the status of the report to "preliminary" and then proceed with the finalization of the report. AGFA issued an Acknowledgment, via FAX- Back that the information was received and understood to affected customers. For any questions customers were instructed to contact their AGFA Healthcare Service at 877-777-2432.
Initiated
2011-06-06
Root cause
Software design
Terminated
2013-07-02
Distribution
Worldwide Distribution - USA (nationwide) and Canada
Recall number
Z-0211-2012
Recall status
Terminated
CDRH record ID
104617
FDA classified
2011-11-18
Recall event number
60154

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗