Open original FDA record ↗Recall record
- FDA identifier
- Z-0211-2015
- Company in FDA record
- Maquet Cardiovascular Us Sales, Llc
- FDA classification date
- Nov 13, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Cardiohelp- i System may have a software issue that can potentially result in an erroneous display of a "Battery Needs Service" message after startup of the units when using either AC or DC power
Source facts
- Action
- Maquet Inc. sent a Product Recall Urgent Medicall Device Field Correction letter to all of their customers via Federal Express. The letter identified the affected product, problem and actions to be taken. The letter stated that a software upgrade will be performed by an authorized Maquet Service Technician. A Maquet representative will contact customer to schedule an on-site service of the device. Any questions can be directed to the local Maquet representative or Maquet Customer Service at 1-888-627-8383 (option 2, option 2) Monday through Friday, between the hours of 8:00 am and 6:00 pm (EST).
- Initiated
- 2014-10-17
- Root cause
- Software Design Change
- Terminated
- 2016-04-06
- Distribution
- Worldwide Distribution - US Nationwide
- Recall number
- Z-0211-2015
- Recall status
- Terminated
- CDRH record ID
- 130939
- FDA classified
- 2014-11-13
- Recall event number
- 69557