Skip to content

Public FDA record

Z-0212-2022: THERAKOS CELLEX Photopheresis Procedural Kit PRODUCT CODE: CLXUSA

Recall record Z-0212-2022 for Therakos, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0212-2022
Company in FDA record
Therakos, Inc.
FDA record creation date
Nov 5, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product was released prior to all testing being completed. If used and a drive tube leak were to occur, may cause illness or injury.

Source facts

Action
Mallincrodt intiated recall by telephone on 10/04/21 and by follow-up letter. Letter states reason for recall, health risk and action to take: If you are receiving this notification, you have been identified as a customer who received CELLEX Procedural Kit Lot #K307. Immediately examine your inventory and quarantine the product subject to this recall. Mallinckrodt Pharmaceuticals will provide you with a pre-paid shipping label to return it to the Mallinckrodt Pharmaceuticals warehouse. If you have any questions regarding the return of the recalled product, please contact Mallinckrodt Pharmaceuticals at (877) 566-9466 to receive further instructions. This recall is being carried out to the direct customer level. Your assistance is appreciated. Please complete and return the enclosed response form as soon as possible.
Initiated
2021-10-04
Root cause
Under Investigation by firm
Distribution
IL
Recall number
Z-0212-2022
Recall status
Open, Classified
CDRH record ID
189725
FDA record created
2021-11-05
Recall event number
88852

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗