Z-0213-2022: Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 7

Recall record Z-0213-2022 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0213-2022
Company in FDA record
Philips North America Llc
FDA record creation date
Nov 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.

Source facts

Action
On September 24, 2021, the firm sent "Urgent Medical Device Correction" recall letters to affected customers (medical facilities). Customers were informed that a software defect causes the possibility for table lock-up and unavailable X-ray imaging when the APC (accept) button and the Float Tabletop (panning) button are pressed or released at the same time. If the problem occurs, there will be a delay in procedure. ACTIONS FOR THE CUSTOMER: - Do not press or release both the APC (accept) button and the Float Tabletop (panning) button at the same time during the procedure. - If the table locks up, do not perform a cold restart. It will take 6 minutes until the system is fully available again. - If the table locks up, perform a warm restart. With a warm restart the system will be fully functional after 90 seconds. Fluoroscopy will be available after 60 seconds. - Place the recall letter with the documentation of the system until Philips has installed the software update in your system. - Return the reply form attached to the recall letter. System restart options are described in section 4.2 of the Instructions for Use. The problem will be resolved by a software update. Customers will be contacted by their local Philips representative to schedule the software update for your system. If you need further information, please contact your local Philips representative: 1-800-722-9377 (reference FCO72200494)
Initiated
2021-09-24
Root cause
Software design
Terminated
2024-09-26
Distribution
Domestic distribution nationwide. International distribution worldwide.
Recall number
Z-0213-2022
Recall status
Terminated
CDRH record ID
189727
FDA record created
2021-11-09
Recall event number
88854

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗