Open original FDA record ↗Recall record
- FDA identifier
- Z-0213-2022
- Company in FDA record
- Philips North America Llc
- FDA record creation date
- Nov 9, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.
Source facts
- Action
- On September 24, 2021, the firm sent "Urgent Medical Device Correction" recall letters to affected customers (medical facilities). Customers were informed that a software defect causes the possibility for table lock-up and unavailable X-ray imaging when the APC (accept) button and the Float Tabletop (panning) button are pressed or released at the same time. If the problem occurs, there will be a delay in procedure. ACTIONS FOR THE CUSTOMER: - Do not press or release both the APC (accept) button and the Float Tabletop (panning) button at the same time during the procedure. - If the table locks up, do not perform a cold restart. It will take 6 minutes until the system is fully available again. - If the table locks up, perform a warm restart. With a warm restart the system will be fully functional after 90 seconds. Fluoroscopy will be available after 60 seconds. - Place the recall letter with the documentation of the system until Philips has installed the software update in your system. - Return the reply form attached to the recall letter. System restart options are described in section 4.2 of the Instructions for Use. The problem will be resolved by a software update. Customers will be contacted by their local Philips representative to schedule the software update for your system. If you need further information, please contact your local Philips representative: 1-800-722-9377 (reference FCO72200494)
- Initiated
- 2021-09-24
- Root cause
- Software design
- Terminated
- 2024-09-26
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
- Recall number
- Z-0213-2022
- Recall status
- Terminated
- CDRH record ID
- 189727
- FDA record created
- 2021-11-09
- Recall event number
- 88854