Skip to content

Public FDA record

Z-0214-2009: Sunquest Laboratory System. Software Version: Sunquest Laboratory v5.4.2, v6.1, v6.2 and v6.3. The sys

Recall record Z-0214-2009 for Sunquest Information Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0214-2009
Company in FDA record
Sunquest Information Systems, Inc.
FDA classification date
Nov 12, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

While performing in-house testing for the Sunquest Laboratory System and under certain conditions, there have been instances in which an error that occurred at the database level or originated in the operating system (Cache)was not communicated or displayed in the Graphic User Interface (GUI) application so the activity continued. While this error was registered in the error log in the System U

Source facts

Action
A notice (PSN-08-07) was issued to all customers on 08/01/2008 via e-mail. The notice advises customers of the problem and actions Sunquest Information Systems has taken to eliminate the Problem. Sunquest Client Support will contact their facility to schedule correcting the error trap setting. The notice also provides temporary procedural instructions to include monitoring error log using System Utilities (Function UTL) and suboption 2, System Error/Message Inquiry. Contact Client Support at 520-570-2252 for questions or additional information.
Initiated
2008-08-01
Root cause
Software design
Terminated
2011-11-28
Distribution
Worldwide Distribution including USA and countries of Canada, Denmark, Saudi Arabia and United Arab Emirates.
Recall number
Z-0214-2009
Recall status
Terminated
CDRH record ID
72843
FDA classified
2008-11-12
Recall event number
49136

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗