Open original FDA record ↗Recall record
- FDA identifier
- Z-0214-2009
- Company in FDA record
- Sunquest Information Systems, Inc.
- FDA classification date
- Nov 12, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
While performing in-house testing for the Sunquest Laboratory System and under certain conditions, there have been instances in which an error that occurred at the database level or originated in the operating system (Cache)was not communicated or displayed in the Graphic User Interface (GUI) application so the activity continued. While this error was registered in the error log in the System U
Source facts
- Action
- A notice (PSN-08-07) was issued to all customers on 08/01/2008 via e-mail. The notice advises customers of the problem and actions Sunquest Information Systems has taken to eliminate the Problem. Sunquest Client Support will contact their facility to schedule correcting the error trap setting. The notice also provides temporary procedural instructions to include monitoring error log using System Utilities (Function UTL) and suboption 2, System Error/Message Inquiry. Contact Client Support at 520-570-2252 for questions or additional information.
- Initiated
- 2008-08-01
- Root cause
- Software design
- Terminated
- 2011-11-28
- Distribution
- Worldwide Distribution including USA and countries of Canada, Denmark, Saudi Arabia and United Arab Emirates.
- Recall number
- Z-0214-2009
- Recall status
- Terminated
- CDRH record ID
- 72843
- FDA classified
- 2008-11-12
- Recall event number
- 49136