Z-0214-2016: Eon 100 system included the following product reference numbers and names: Reference no. 77345- Glucose

Recall record Z-0214-2016 for ELITech Clinical Systems SAS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0214-2016
Company in FDA record
ELITech Clinical Systems SAS
FDA classification date
Nov 5, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ELITech clinical Systems reagents based on Trinder reaction used on the Eon 100 instrument system are recalled due to potential drug interference. No complaints were received.

Source facts

Action
The firm sent recall letters or Technical Bulletins to US consignees on 9/22/15 via registered mail. Technical Bulletin #1022 (for EON 100 users) Technical Bulletin #2008 (for EON 300 users) Technical Bulletin #5064 (for ENVOY users) Technical Bulletin #3009 (for Selectra users) ELITechGroup is not requesting return of the affected product. The method sheets and package inserts will be updated with the new interference claims. When available, the revised labeling will be provided to the customer for customer's use until the new IFU is included in the product packaging. Consignees are requested to return the Confirmation of Receipt form back to the firm by FAX to 401-642-9001 or send email to customercare@elitechgroup.com or send via US Post: Vital Diagnostics, 27 Wellington Rd., Lincoln, Rl. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822 or 855-ELITECH
Initiated
2015-09-22
Root cause
Under Investigation by firm
Terminated
2016-03-04
Distribution
US Nationwide Distribution
Recall number
Z-0214-2016
Recall status
Terminated
CDRH record ID
140616
FDA classified
2015-11-05
Recall event number
72322

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗