An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Source facts
Action
On September 19, 2019 the firm distributed Urgent Field Safety Notice (FSN) Medical Device Correction letters by direct communication from Philips Field Service Engineers instructing them to:
- Discontinue system use until further notice
- Inform those who need to be aware within your organization or any organization where the potentially affected devices have been transferred (If appropriate).
- Maintain this notice with your system Instructions for Use (IFU) until the correction is made to the system.
ACTIONS PLANNED BY PHILIPS: Philips Healthcare is distributing this FSN to all affected customers/users and will deploy a solution addressing the issue upon completion of the investigation.
FURTHER INFORMATION AND SUPPORT: If you need any further information or support concerning this issue, please contact your local Philips representative: For SUPPORT in North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts).
Initiated
2019-09-19
Root cause
Nonconforming Material/Component
Terminated
2021-04-05
Distribution
Nationwide Distribution.
Recall number
Z-0216-2020
Recall status
Terminated
CDRH record ID
177006
FDA classified
2019-11-01
Recall event number
84009
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.