Z-0216-2026: DxI 9000 Access Immunoassay Analyzer, Part Number C11137

Recall record Z-0216-2026 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0216-2026
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Oct 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 9/12/25 was sent to customers. " When ordering tests for a manually diluted sample through the Test Order Entry page, a unique sample ID is required. " Alternatively, use the Dilute and Rerun feature or an automated dilution to measure diluted samples. Refer to Rerunning a Test with a Diluted Sample in the System Help for more information. " Beckman Coulter suggests sharing this letter with your Medical Director to determine if performing a retrospective review of results is necessary. " Beckman Coulter is investigating the root cause of this issue and will implement a correction with a future software release. " Your Beckman Coulter service representative will contact you to schedule the software upgrade when it is available. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact Customer Support Center: " From our website: http://www.beckmancoulter.com
Initiated
2025-09-19
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-0216-2026
Recall status
Open, Classified
CDRH record ID
216201
FDA classified
2025-10-21
Recall event number
97638

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗