Z-0217-05: IMx HAVAB Controls are for In Vitro Diagnostic use. The kit is composed of 2 bottles of Negative and Posi

Recall record Z-0217-05 for Abbott Health Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0217-05
Company in FDA record
Abbott Health Products, Inc.
FDA classification date
Nov 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Abbott identified through investigational studies that IMx HAVAB Controls lot 18220Q100 is generating Negative Control values outside the upper range specified in the IMx HAVAB Package Insert.

Source facts

Action
Device recall letters were dated 10/27/2004. Letter was mail to customers asking to destroy and discontinue use of device. On hand inventory information was requested to be forward to Abbott at 1-800-777-0051 (U.S. only).
Initiated
2004-10-27
Root cause
Other
Terminated
2005-03-16
Distribution
The product has been distributed to hospitals, laboratories and other institution including government accounts. Geographic areas of distribution include Puerto Rico, Canada, Germany, Thailand and U.S.A. Government accounts includes: VA Medical Center, Saint Louise, MO; Brecksville VA Hosp. Supp, Brecksville, OH; VA MED CENTER, Phoenix, AZ; Department Veterinary Affair Med., Houston, TX; VA Med., Cincinnati, OH; VA Med Center, Seattle, WA; Truman Mem. VA, Columbia, MO; Dept. Vet. Aff., Shreveport, LA; VA Hosp. Whse., Fresno, CA ; VA Med Center, Omaha, NE ; VA Med. Center, Lexington, KY; VA Med Center, Detroit, MI ; VA Hosp., San Diego, CA ; Naval Hosp., Great Lakes, IL; VA Med. Center, San Francisco, CA.
Recall number
Z-0217-05
Recall status
Terminated
CDRH record ID
35782
FDA classified
2004-11-20
Recall event number
30388

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗