When the device is subject to physical shock, such as a drop to the floor, it may output a different value from the preset value. This may result in a superficial burn on the patient if the stimulation is done for a long time.
Source facts
Action
All customers will be called by phone, then faxed the Product Notification starting 09/01/2006
Customers were instructed to indicate the serial numbers on each of your MS-210BK stimulators and fax the MS-210BK PRODUCT RECALLREPLACEMENT letter back to Nihon Kohden at: 949-580-15550 or email to customerservice@nkusa.com. Upon receipt of this letter, Nihon Kohden''s customer service department will send a replacement unit.
Initiated
2006-08-23
Root cause
Other
Terminated
2008-05-23
Distribution
Nationwide.
Recall number
Z-0218-2007
Recall status
Terminated
CDRH record ID
48578
FDA classified
2006-11-28
Recall event number
36394